39
f) Analog Input Calibration
NOTE: The analog inputs available for the TIO were calibrated at the
factory. There is no need to perform analog input calibration unless
the CPU IC, input amplifier IC or passive components from the analog
input circuitries were replaced. Any alteration of the analog input
scaling variables in the EEPROM table will VOID the calibration
warranty supplied with instrument.
,
NOTE: Check the actual input jumpers’ configuration before applying
any input signal to TIO. Be sure your input signal does not exceed
the maximum allowed level for corresponding input type (see Table
4.1). Do not apply voltages above 5.0 Vdc unless TIO input was
specifically configured in the factory for 0-10 Vdc (check actual
model number and specification). Exceeding maximum allowed input
level may cause inadvertent damage to the device circuitry.
,
The TIO analog output calibration involves calculation and storing the off-
set and span variables in the EEPROM based on two calibration points (0
and 100% F.S.). The 0-5 (0-10)Vdc output has only scale variable and 4-
20 mA output has offset and scale variables.
Power up the TIO instrument for at least 15 minutes prior to commenc-
ing the calibration procedure. Observe analog input jumper position (see
Figure 4.2) and connect corresponding type of the calibration signal
source device to pins 4 (plus) and 8 (minus) of the 9-pin D-connector.
Follow firmware prompts and apply calibration point values according to
on screen instructions. If calibration must be aborted, press ESC but
ton. When calibration is completed firmware will display new offset and
span values and ask press ENT button to save new calibration variables
to EEPROM or ESC to abort calibration and exit without saving. In the
end, the firmware will prompt with a confirmation message.
g) Pilot Calibration Timer
The Pilot Calibration timer accumulates operational hours since the last
time the unit was calibrated. The smallest increment value is 0.1 Hour
(6 minutes). The value of the timer may be reset by the user by pressing
RIGHT button. Once RIGHT button is pressed the confirmation screen
will appear with the “Yes” or “No” menu. Selecting “Yes” will reset the
pilot calibration timer back to zero.