Aalborg TIO Totalizer I/O Flow Monitor/Controller Manuel d'utilisateur Page 59

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 98
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 58
56
TABLE 6.2 J2 INPUT / OUTPUT JUMPER CONFIGURATION OPTIONS
FOR GFM SERIES FLOW METERS
PV INPUT
TYPE
(TIO INPUT)
J2 JUMPER CONFIGURATION
GFM
CABLE KIT
NOTE
J2A J2B J2C J2D J2E J2F
0 - 5 Vdc 2 - 3 5 - 6 8 - 9 10-11 14-15 17-18
KIT-TM-DD
KIT-TM-FD
5 - 10 Vdc 2 - 3 5 - 6 8 - 9 11-12 14-15 17-18 KIT-TM-RD
+5Vdc
reference
signal must
be used
0 - 10 Vdc 2 - 3 5 - 6 8 - 9 11-12 14-15 17-18
NOT
SUPPORTED
FOR GFM
Special Order
option! (PCB
hardware must
be changed)
4 – 20 mA 2 - 3 5 - 6 8 - 9 10-11 13-14 16-17
N/A (user
custom cable
assembly)
(249 Ohm
passive, not
isolated current
input)
d) Parameters Configuration
Following parameters must be configured:
Device Function (See 4.3.12 Submenu “Device Function”). “Meter”
function must be selected.
Full Scale Range (See 4.3.14 Submenu “Device Calibration”). Full
Scale Range parameter must be set equal to the GFM full scale flow
rate in Litr/min.
Fluid Std. Density (See 4.3.14 Submenu “Device Calibration”). This
parameter is required only when mass-based engineering units are
selected.
User may configure other parameters (see Paragraph 5.3) according
to individual preferences and application requirements.
NOTE: If “Full Scale Range”, “Device Function” and “Fluid Std.
Density” parameters are not set properly the device may have
erroneous reading and unpredictable behavior.
,
Vue de la page 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 97 98

Commentaires sur ces manuels

Pas de commentaire