Flow Alarms
Sets / reads the parameters of
the Flow Alarms.
Note: If device function is set
to “Meter” the High Alarm
value must be more than Low
Alarm value.
Meter Alarm conditions:
Flow ≥ High Limit = H
Flow ≤ Low Limit = L
Low < Flow < High = N
Controller Alarm conditions:
Flow-SP High Limit = H
SP-Flow Low Limit = L
Low > SP-Flow < High = N
8 A
C
(set Alarm
configuration)
<Value>
(high limit, %F.S.)
<Value>
(low limit, %F.S.)
AC:40.1,20.5
A
(action delay in
sec.)
<Value>
[0-3600]
AA:<Value (sec)>
E
(enable Alarm)
A:E
D
(disable Alarm)*
A:D
R (read
current status)
AR:N (no Alarm)
AR:H (High Alarm)
AR:L (Low Alarm)
S (Read
current settings)
AS:M,H,L,D,B
where: M – mode (E/D)
H – High settings value
L – Low settings value
D – Action Delay (sec)
B – Latch mode (0-1)
Example:
AS:E,40.0,20.0,2,0
L
(Latch mode)
<Value>
(0-disabled*)
(1-enabl’d)
AL:<Value> where:
Value = 0 – 1
Example:
AL:0