Aalborg TIO Totalizer I/O Flow Monitor/Controller Manuel d'utilisateur Page 73

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 98
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 72
70
NO. INDICATION LIKELY REASON SOLUTION
5
Erratic Flow rate
Reading
Wrong configuration of
J2 Input/Output
Jumpers
Check J2 input jumper configuration (see
Table 6.4) If necessary contact factory for
additional help.
TIO “Full Scale Flow”
parameter value (in
litr/min) is not equal to
the mated device full
scale range
Check settings for “Full Scale Range (See
Paragraph 5.3.14 Submenu “Device
Calibration”). Full Scale Range parameter
must be set equal to the mated device full
scale flow rate in Litr/min.
TIO “Fluid Std. Density”
parameter is not set
according to fluid being
used and mass-based
engineering units are
selected.
Check settings for “Fluid Std. Density” (See
Paragraph 5.3.14 Submenu “Device
Calibration”). This parameter is required
only when mass-based engineering units are
selected.
6
Totalizer reading is
wrong
Wrong configuration of
J2 Input/Output
Jumpers
Check J2 input jumper configuration (see
Table 6.4) If necessary contact factory for
additional help.
TIO “Full Scale Flow”
parameter value (in
litr/min) is not equal to
the mated device full
scale range
Check settings for “Full Scale Range (See
Paragraph 5.3.14 Submenu “Device
Calibration”). Full Scale Range parameter
must be set equal to the mated device full
scale flow rate in litr/min.
7
LCD Displays flow read-
ing, but communication
interface does not work.
Wrong host PC inter-
face, or wiring connec-
tion.
Make sure interface type (RS-232 or RS-
485) on the host PC is the same as on TIO
device. If required, use RS-232 to RS-485
converter. Check communication wiring
connection according to Paragraph 4.4
TIO has RS-485 inter-
face, butt device
address does not match
to address used by host
PC.
Change TIO RS-485 address to be matched
to host PC software settings (see Paragraph
5.3.13 b).
8
The Device Diagnostic
Alarm Event with code
0 – “CPU Temp. High”
is active.
MCU / PCB temperature
is too high (overload).
Disconnect power from the TIO. Make sure
the ambient temperature is within specified
range (below 70 °C). Let the device cool
down for at least 15 minutes. Apply power
to the device and check Diagnostic Alarm
Event. If overload condition will be indicated
again the unit must be returned to the facto-
ry for repair.
9
The Device System
Event with code D -
“Fatal Error” is active.
Fatal Error (EEPROM or
SRAM corrupted)
Cycle the power on the TIO. If System Event
with code D indication again the unit must
be returned to the factory for repair.