COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3 ARGUMENT 4 RESPONSE
Pulse
Output
Sets and controls action of the
programmable Pulse Output
circuitry. NOTE: Unit/Pulse
value has to be entered:-for
Volumetric flow conditions
in currently selected volume EU
-for Mass flow conditions
in currently selected mass EU
EU has to be not time based.
It is recommended to set the
unit/pulse value equal to the
maximum flow in the same
units per second. This will limit
the pulse to no faster than one
pulse every second.
Example: Maximum flow rate:
1200 liter/min (1200 liter/min =
20 liters per second)
If Unit/Pulse is set to 20 liters
per pulse, the output will pulse
once every second.
13 P
C (flow
configuration)
V-volum. flow
M-mass flow
<Value>
(Unit/Pulse)
Volumetric flow -
current
Volumetric E.U.
Mass flow - cur-
rent mass EU
(example:
10 litr/pulse)
PCV:<value>,<value>
Example:
PCV:V,10
Or
PCM:<value>,<value>
Example:
PCM:M,2.5
T (Pulse active
Time in ms)
<value>
[10 -13000 ms]
PT:<value>
Example: PT:1000
D (disable pulse
output)*
P:D
E (enable pulse
output)
P:E
Q (read current
pulse output
Queue value)
PQ:<value>
(number of pulses)
S (setting status) PS:Mode,FlowConditions,
Unit/Pulse,PulseTimeInterval
Fluid
Density for
standard
conditions
Read and set Fluid Density for
standard conditions in g/cm3
NOTE: it is not measured
density.
14 FD
(New Density
Value)
0.0001 Density
10.0 g/cm3
FD:<value>
Example:
FD:1.293
No Argument
(Returns Current
Density in g/cm3)
FD:<value>
Example:
FD:1.293